For full functionality of this site it is necessary to enable JavaScript.

HIOKI LR8400-20 HiLOGGER MEMORY

SPECIFICATIONS
- Analog input UNIVERSAL UNIT LR8501:
   + 15 analog channels
   + Voltage:10mV~100V, 1-5V f.s. 10 ranges, resolution: 500nV
   + Temperature: -200°C to 2000°C. resolution 0.01°C
   + Temperature: Pt 100, JPt 100 sensor. -200°C~800°C, resolution 0.01°C
   + Resistance: 10 Ω~200 Ω f.s.
   + Humidity: 100 % rh f.s. 5.0~95.0 % rh/ resolution 0.1% rh
- Analog input VOLTAGE/TEMP UNIT LR8500:
   + 15 analog channels
   + Voltage: 10 mV~100 V, 1-5 V f.s. / resolution: 500 nV
   + Thermocouples: -200°C~2000°C / resolution 0.01°C
   + Humidity 100 % rh f.s. 5.0~95.0 % rh / resolution 0.1 % rh
- Pulse, Digital input:
   + 8 channels, pulse / digital
   + Pulse totalization: 0~1000 M pulse
   + Pulse revolutions: 0 to 5000 /n (r/s) f.s/ resolution 1/n(r/s)
- Recording intervals: 10ms~50ms, 100ms~1hour, 19selections
- Digital filter: Select from OFF/ 50 Hz/ 60 Hz
- Data storage: 8 M-words, CF card or USB memory
- Inteface: LAN-100BASE-TX/ USB2.0
- Display device: 5.7 inch TFT color liquid crystal display
- Other functions: Save data CF card or USB memory stick
- Power supply:
   + AC ADAPTER 9418-15: 100 to 240 V AC, 50/60 Hz), 7VA
   + BATTERY PACK Z1000
- Product Net Weight: 1.8 kg
- Product Size: 272*182*66.5mm
- Supplied accessories:
   + Detailed operating manual ×1, Measurement guide ×1
   + AC ADAPTER 9418-15 ×1, USB cable ×1
   + CD-R (data collection software "Logger Utility") ×1
- The warranty period: 12 month
Details

 

Analog input UNIVERSAL UNIT LR8501
Actual functionality will depend on combination of units installed

Note: Isolated from each channel to chassis
[No. of channels] 15 analog channels; isolated scanning method input (4 terminals: push-button type)
[Voltage] 10 mV to 100 V, 1-5 V f.s. 10 ranges, Max. resolution: 500 nV
Note: Isolated between channels and from each channel to chassis
[Temperature: Thermocouples] -200°C to 2000°C (depends on sensor), 3 ranges (K, J, E, T, N, R, S, B, W), Max. resolution 0.01°C
Note: Isolated between channels and from each channel to chassis
[Temperature: Pt 100, JPt 100 sensor] -200°C to 800°C, 3 ranges, Max. resolution 0.01°C
Note: Not isolated between channels
[Resistance] 10 Ω to 200 Ω f.s. 4 ranges, Max. resolution 0.5 mΩ
Note: Not isolated between channels
[Humidity] 100 % rh f.s. 5.0 to 95.0 % rh (when using Z2000), resolution 0.1 % rh
Note: Not isolated between channels nor from each channel to chassis
[Max. rated voltage between isolated input channels] 300 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC
Analog input VOLTAGE/TEMP UNIT LR8500
Actual functionality will depend on combination of units installed

Note: Isolated from each channel to chassis
[No. of channels] 15 analog channels; isolated scanning method input (2 terminals: M3 screw type)
[Voltage] 10 mV to 100 V, 1-5 V f.s. 10 ranges, Max. resolution: 500 nV
Note: Isolated between channels and from each channel to chassies
[Temperature: Thermocouples] -200°C to 2000°C (depends on sensor), 3 ranges (K, J, E, T, N, R, S, B, W), Max. resolution 0.01°C
Note: Isolated between channels and from each channel to chassies
[Temperature: Pt 100, JPt 100 sensor] [Resistance] not available
[Humidity] 100 % rh f.s. 5.0 to 95.0 % rh (when using Z2000), resolution 0.1 % rh
Note: Not isolated between channels nor from each channel to chassies
[Max. rated voltage between isolated input channels] 250 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC
Caution: Max. voltage from terminals to ground without damage
Pulse, Digital input [No. of channels] 8 channels, pulse / digital selectable for each channel, M3 screw terminal, not isolated, common ground
[Pulse totalization] 0 to 1000 M pulse, 1 range (No-voltage a contact; normally open, open collector or voltage input), Max. resolution 1 pulse
[Pulse revolutions] 0 to 5000 /n (r/s) f.s. 1 range (same as Pulse totalization input signal condition), resolution 1/n (r/s) Note: "n" is the number of sensor output pulses per revolution, 1 to 1000
[Digital input] Record logical 1 or 0at each sampling
[Max. rated voltage between input channels] Not isolated
[Max. allowable input] 0 to 50 V
Recording intervals 10 ms to 50 ms, 100 ms to 1 hour, 19 selections (All input channels are scanned at high speed during every recording interval)
Note: limited by using channels at 10 ms to 50 ms interval
Digital filter Select from OFF/ 50 Hz/ 60 Hz (the cut-off frequency is automatically set)
Data storage Internal memory: 8 M-words, Data storage media: CF card or USB memory
(Only data recorded to a genuine HIOKI CF card is guaranteed)
LAN interface 100BASE-TX, Functions: Data acquisition using bundled software or PC commands, FTP server, FTP client, HTTP server function, or E-mail system
Note: LAN communication support planned from software Ver. 1.20
USB interface USB 2.0 High-speed capable, series mini-B receptacle
Functions: Data acquisition using bundled software or PC commands, Transfer data from the CF card to a PC via USB drive mode
Note: Data transfer not possible from USB memory sticks
Display device 5.7 inch TFT color liquid crystal display (640 × 480 pixel)
Other functions Save waveform data in real time to the CF card or USB memory stick, Numerical value calculations, Waveform calculations, and others
AC Power Using the AC ADAPTER 9418-15 (100 to 240 V AC, 50/60 Hz), 7 VA
DC Power Using the BATTERY PACK Z1000 (optional accessory), Continuous use 5 hr, External power: 10 to 28 V DC (Please contact your HIOKI distributor for connection cord)
Dimensions and Mass 272 mm (10.71 in) W × 182.4 mm (7.18 in) H × 66.5 mm (2.62 in) D, 1.8 kg (63.5 oz), (LR8400-20 main unit, excluding the Battery Pack 370 g/ 13.1 oz)
Accessories Detailed operating manual ×1, Measurement guide ×1, AC ADAPTER 9418-15 ×1, USB cable ×1, CD-R (data collection software Logger Utility) ×1

Catalog

  • Quality Engagement
  • Easy change and return
  • Delivery Avaliable
  • Favorable payment

Apply your mail to get promotion information